EasyManua.ls Logo

Protec PRS 500 E - 4.12 Automatic Exposure Control

Protec PRS 500 E
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PRS 500 E
Instructions for use
5069-0-8002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 29 von 47
4.12 Automatic Exposure Control
If the PRS 500 E is operated with an automatic exposure control, the functionality can be checked as
follows:
Place a phantom or any other weakening object (not lead) in the beam path. Select the measurement
chamber field to be checked and trigger the exposure. If this works properly, the measured value is
displayed. Repeat this procedure for all existing measuring chamber fields. If an error message is
displayed during this test, please contact PROTEC service department or a service company authorized
by them immediately.

Table of Contents

Other manuals for Protec PRS 500 E

Related product manuals