PULSAR MEASUREMENT
21
Rack and Panel
The Ultra 3 rack mount unit is a standard 10HP plug in unit, for a standard
84HP (19”), 160 mm sub rack unit.
The connector that is supplied with the rack unit should first be fitted to the
back of the sub rack, so that when the rack unit is inserted into the rack, it
connects correctly. To install the unit into the sub rack, slide the Ultra 3 into
the rack and secure by tightening the four-captivated screw that are in the
faceplate.
The panel mount unit should be installed by cutting a hole of size DIN 68
+0.7
x 138
+1.0
mm in the panel, then insert the Ultra 3 enclosure through the hole,
and tighten the fixing screws from the rear. Make sure you leave sufficient
room for the cables behind the enclosure (a depth of at least 225 mm
behind the panel should suffice).
Dimensions of the rack enclosure are as shown below:
172.5mm
188 mm
128.5mm
10 HP (50.4mm)
35.6mm
VOLTAGE SELECTOR
230
115
AC POWER FUSE