EasyManua.ls Logo

Pulsar Measurement Ultra 3 - U-Throated Flume Calculations

Pulsar Measurement Ultra 3
179 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PULSAR MEASUREMENT
85
U-Throated Flume Calculations
Absolute
If the flow calculation is to be absolute P702 = 1 the flow will be calculated
using the formula: q = (2/3)
1.5
g
n
0.5
C
u
C
v
C
d
bh
1.5
Where:
q = flowrate
g
n
= gravitational acceleration, (nominal value = 980.66 cm/s
2
)
h = head
C
u
= shape coefficient calculated by Ultra 3 (P724)
C
v
= velocity coefficient calculated by Ultra 3 (P721)
C
d
= discharge coefficient calculated by Ultra 3 (P722)
b = throat width P711
Ratiometric
If the flow calculation is to be ratiometric P702 = 2 the flow will be
calculated using the formula:
q= q
cal
(C
v
/C
vcal
)(C
d
/C
dcal
)(C
u
/C
ucal
)(h/h
cal
)
1.5
Where:
q = flowrate
q
cal
= flowrate at maximum head P705
Cv = velocity coefficient calculated by Ultra 3 (P721)
Cv
cal
= velocity coefficient at maximum head
Cd = discharge coefficient calculated by Ultra 3 (P722)
Cd
cal
= discharge coefficient at maximum head
Cu = shape coefficient P724
Cu
cal
= shape coefficient at maximum head
h = head h
cal
= maximum head P704

Table of Contents