PULSAR MEASUREMENT 
 
15  
 
CALIBRATION 
 
Press  or  to position curser at Calibration menu, and  to enter. Use  
or  to position cursor before each menu item and  to enter. When settings 
are completed press  to store and  again to return to the Main Menu. 
 
20mA 
(5V)    
Press  then  or  to change the numbers and 
decimal point. Use this menu to set the corresponding 
flow rate that will be represented by 20mA analog 
output. If maximum flow is unknown, enter an estimated 
flow rate and observe actual flow to determine the 
correct maximum value. Any velocity or flow rate up to 
+40 ft/sec (12.2 m/sec) may be selected. 
 
4mA 
(0V) 
Press  or  to set the flow rate corresponding to 4mA 
analog output. This setting may be left at zero flow (or 
velocity or can be raised to any value less than the 
20mA setting, or lowered to any velocity or 
corresponding flow rate down to -40 ft/sec (-12.2 
m/sec). 
 
Min Flow 
Press  and enter a minimum flow cutoff. Forward and 
reverse flows less than   Min Flow   will be forced to 
zero. 
 
Signal Cutoff 
 
 
 
Adjust the setting in percent to suppress flow readings 
at zero flow when fluid swirling or pipe vibration may 
cause the instrument to continue reading.  Example: 
Signal Cutoff at 5% will force the display and outputs to 
zero when signal strength drops below 5%. 
 
LOS Time 
 
Selectable from 0-99 seconds, default of 10, the LOS 
Time controls how long the Signal Strength must be 
below the Signal Cutoff before the reading on the DFM 
6.1 goes to zero. This parameter is used to suppress 
sudden losses of signal which might be caused by a 
partially empty pipe, or cleaner fluids flowing by the 
sensor.  
 
 
 
   
 
 
   
       
- - Ca l i b r a t i o n - - - - - - -
 Mo d e Fl o w 
   2 0 mA   5 0 0  USG/ m
      4 mA 0  USG/ m
  M   5 . 0 0  USG/ mi n  Fl o w
   Si g n a l  Cu t o f f
    Da mp i n g
      Mo d e FI R
     Pe r c e n t 1 0 %
      Wi n d o w 1 . 0  f t / s
   Pi p e  I D 4 . 0 3  i n
   Ca l  Co n s t a n t 1 . 0 0 0