DFM 6.1 INSTRUCTION MANUAL
12
24 HR LOG
Press from the main display to view a formatted flow report from
instruments with a built-in data logger. Press to scroll down one day or
repeatedly to scroll to a specific date. Up to 365 days can be stored. Newest
date will overwrite the oldest. Press to return to the main display.
IMPORTANT: Inserting a USB drive into the meter while on this screen will
transfer 24 HR Log data to the USB drive in .csv format.
PASSWORD
The Password (a number from 0000 to 9999) prevents unauthorized access to
the Calibration menu.
From the Main Display press the key to get to Password. Factory default
password is 0000 and if it has not been changed, this screen will be bypassed
completely.
A new password can be stored by going to the Special Functions New
Password menu.
If a user password is required, press to place the cursor under the first digit
and or to set the number, then to the second digit, etc. Press or
to proceed to the Menu Selections screen.
MENU SELECTIONS
The Menu selections page is used to navigate to specific menus which are
described in more detail on the following pages.
Press or to navigate to different menus, and to enter the selected
menu.
- - Me n u - - - - - - - - - - - - - -
Un i t s / Mo d e
Ca l i b r a t i o n
Re l a y Pa r a me t e r s
Da t a L o g g i n g
Co mmu n i c a t i o n
Sp e c i a l F u n c t i o n s
Si mu l a t i o n
Co n f i g u r a t i o n