Page 146
Vessel Shape
P600 Value
Dimensions
Required
P600=4 Cylindrical
Parabola base
Cylinder diameter and
height of bottom
P600=5 Cylindrical
Half-sphere base
Cylinder Diameter
P600=6 Cylindrical
Flat sloped base
Cylinder diameter and
height of bottom
P600=7 Rectangular
Flat sloped base
Width and Breadth of
rectangular section and
height of bottom
P600=8 Horizontal
cylinder with flat ends
Cylinder diameter and
tank length
P600=9 Horizontal
cylinder with
parabolic ends
Cylinder diameter,
length of one end
section, and tank
length
P600=10 Sphere
Sphere diameter
P600=11 Universal
Linear
No dimensions
required, level and
volume breakpoints
used.
P600=12 Universal
Curved
No dimensions
required, level and
volume breakpoints
used.