Page 154 
P707 Time Units 
Select the Time Units to be used with the Volume Units to determine the 
desired flow rate from the options below: 
 
P708 Flow Decimal 
This parameter determines the  number  of decimal places  in the  flow rate 
reading during run mode. It can be set between 1 and 3. Default = 2 
P709 Flow Cut Off 
This  parameter is  used  to  select  the  minimum flow,  in  a  %  of  flow  rate, 
which is to be totalised. Enter values in % of maximum flow. Default = 5% 
 
Dimensions 
 P710 Dimension A 
This parameter is used to enter dimension “A” of the Primary Measuring 
Device, where applicable, see table below for further details. 
P711 Dimension B 
This parameter is used to enter to enter dimension “B” of the Primary 
Measuring Device, where applicable, see table below for further details. 
P712 Dimension C 
This parameter is used to enter to enter dimension “C” of the Primary 
Measuring Device, where applicable, see table below for further details.