EasyManua.ls Logo

Pulsar Ultra 3 (UL) - Dimensions

Pulsar Ultra 3 (UL)
209 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 154
P707 Time Units
Select the Time Units to be used with the Volume Units to determine the
desired flow rate from the options below:
Option
Description
1= per Second
Flowrate will be calculated and displayed in
Volume units/Second
2= per Minute
Flowrate will be calculated and displayed in
Volume units/Minute
3= per Hour
Flowrate will be calculated and displayed in
Volume units/Hour
4= per Day (Default)
Flowrate will be calculated and displayed in
Volume units/Day
P708 Flow Decimal
This parameter determines the number of decimal places in the flow rate
reading during run mode. It can be set between 1 and 3. Default = 2
P709 Flow Cut Off
This parameter is used to select the minimum flow, in a % of flow rate,
which is to be totalised. Enter values in % of maximum flow. Default = 5%
Dimensions
P710 Dimension A
This parameter is used to enter dimension “A” of the Primary Measuring
Device, where applicable, see table below for further details.
P711 Dimension B
This parameter is used to enter to enter dimension “B” of the Primary
Measuring Device, where applicable, see table below for further details.
P712 Dimension C
This parameter is used to enter to enter dimension “C” of the Primary
Measuring Device, where applicable, see table below for further details.

Table of Contents

Related product manuals