Page 172
P904 Peak Noise1
This is the peak noise reading for the transducer. It is measured while the
transducer is not firing, and gives an indication of the maximum amount of
electrical noise present on the cabling.
P905 Sensitivity
This parameter determines the sensitivity of the unit. Please consult Pulsar
for further information and assistance on changing the value of this
parameter.
P906 Side Clearance
This parameter is used to set the distance by which the DATEM trace will
stand-off from around unwanted echoes such as obstructions. Please
consult Pulsar for further information and assistance on changing the value
of this parameter.