4 65000159 Rev. 008 (06/17)
FastPack
®
IP System Set Up
Getting Power to the FastPack
®
IP System
Action
Instrument
Information
Illustration
Plug Desktop Power Supply
into the analyzer.
Plug AC Power Cord into
Desktop Power Supply.
Turning ON the System Analyzer
Plug AC Power Cord into
the wall socket.
Turn analyzer on using
power button located on
back of unit.
Attention: Leave analyzer
on at all times.
Power Supply operates
with input voltage of 100
240±10% VAC and 50/60
Hz.
Software vX.X
The analyzer will beep
once and display the
internal software version
for 5 seconds. Analyzer will
now perform a self-
diagnostic check and it will
not function until
operating temperature is
reached. It takes 40- 60
minutes to reach operating
temperature. During this
time the display will show
two sets of temperature
readings. The analyzer will
beep twice and display
“Ready” when operating
temperature is reached.
Ready
6/21/15
Ready
6/21/15