Analyzer Software Upgrade Using USB
1. Power off the analyzer
2. Insert USB drive into the USB port
in back of the analyzer
3. Power on the analyzer.
The screen will be blank for about 15
seconds.
The newly installed software version
number will then appear for 2
seconds.
The Analyzer will automatically
perform self-diagnostics for about 3
minutes, and then the screen will
display “Ready”.
4. Confirm the software was loaded
successfully.
Note: If the software was not
loaded successfully, repeat all
previous steps
With “Ready” on the screen, press
the red button until the Software
Version Number appears.
Confirm that the software version
number is the version you
intended to install.
Software
VX.XX
Ready