65000159 Rev. 008 (06/17) 23
Control Testing Procedure
Action
Instrument
Information
Illustration
Closely follow FastSteps 2
through 5 from your
FastSteps Guide.
Utilize only Qualigen
recommended pipette tips.
The tips are available
through your local
distributor or Qualigen.
Be sure that the pipette
operates smoothly and
does not stick.
It is very important that
the tip is properly seated
on the pipette. If it is not,
there is a high probability
of short sampling which
will directly affect control
results.
Closely follow FastSteps 6
through 10 from your
FastSteps Guide.
Use serum, plasma,
calibrators or controls
only.
Always wear gloves when
handling samples.
It is very important that
the pipette tip is firmly
placed in the injection
port. If any sample leaking
is detected, dispose of the
FastPack
®
IP and begin
again with a new FastPack
®
IP and sample.
FastPack
®
IP Vitamin D Users ONLY! Sample Pretreatment
Eject the control sample
into the pre-treatment
buffer tube by pressing
down on the pipette
plunger to the first stop.
Replace the screw cap
tightly onto the buffer
tube.
Each patient sample,
calibrator or control must
be thoroughly mixed with
the FastPack
®
Pre-
treatment buffer before
running a FastPack
®
IP
Vitamin D assay.