65000159 Rev. 008 (06/17) 55
Cleaning the FastPack
®
IP System
Note: Always wear protective gloves when cleaning or
disinfecting the analyzer.
The FastPack
®
IP System analyzer is completely self-
enclosed, and only requires periodic cleaning to remove
excess sample. Use a damp cloth to wipe down the entire
exterior and interior compartment of the analyzer. Do
not use solvents to wipe down the exterior of the
analyzer.
If a FastPack
®
IP has leaked inside the FastPack
®
IP
analyzer, perform the following cleaning procedure:
FastPack
®
IP System Cleaning Procedure
Wipe the inside of the door panel with a soft cloth soaked with a mild disinfectant or
any hospital grade disinfectant to remove any potential contaminant that may
impact performance of the analyzer.
Note: Do not press too hard as the silicone membrane can be torn with excessive
force.
Wipe the membrane side of the analyzer chamber with a mild disinfectant to
eliminate any potential active chemiluminescence chemicals that may impact
performance.