12 65000159 Rev. 008 (06/17)
Calibration Procedure
Action
Instrument
Information
Illustration
Open the FastPack
®
IP
Calibration Kit. Remove
contents.
Be sure to refer to the
Direction Insert for the test
type being calibrated.
Important quality control
information is now
included in your Quality
Assurance Log.
Place “Calibration card” on
the door. Align the pins on
the door with the holes in
the FastPack
®
IP.
The analyzer will read the
bar code on this card to
perform the calibration. If
the calibrator has expired,
the analyzer will not
operate.
Press the BLUE button.
Ready
The door will lock, and the
analyzer will read the
calibration card. This will
take about 30 seconds.
Expiration Date
Test Type
Calibrator Lot #
Calibration Kit
Lot Number