22 65000159 Rev. 008 (06/17)
FOR FASTPACK
®
IP VITAMIN D USERS ONLY
8. Pretreatment Buffer
Control Testing Procedure
Action
Instrument
Information
Illustration
Open the FastPack
®
IP
Control Kit. Remove
contents.
Keep the Control Range
Card in a place easily
retrieved for reference;
possibly in your Quality
Assurance Log.
Write “CX”, depending
upon the Control Vial used
and lot number, and the
operator’s initials on the
peel-off FastPack
®
IP label.
E.g., C1 for Control Level 1
or C2 for Control Level 2.
Invert Control Vial 6-8
times before opening.

Table of Contents