65000159 Rev. 008 (06/17) 29
Sample Dilution - hCG
When to Dilute a Test Sample
Samples can be accurately measured for hCG if it is within the analytic range of 1. 8 – 1000 mIU/mL.
However, in some instances, the analyte level in patient samples may exceed the assay range of 1000
mIU/mL. In these cases, the assay must be run again using a diluted sample in order to determine the
actual quantitative hCG concentration.
What You Need
1. 25 µL Pipette
2. Pipette Tips
3. Patient Sample
4. FastPack
®
hCG Sample Diluent Kit
5. FastPack IP for hCG
6. Microcentrifuge
7. Medical Gloves – used to avoid contact with
biohazard materials such as human blood, etc.