EasyManua.ls Logo

Qualigen FastPack IP System - Calibration Reset; Calibration Reset Procedure

Default Icon
87 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
65000159 Rev. 008 (06/17) 67
Calibration Reset
Calibration Reset Procedure
Action
Instrument
Information
Illustration
Ready
Place the Calibrator Card,
for the test type
calibration you want to
erase, on the door. Align
the pins on the door with
the holes in the Calibration
Card.
Ready
Close the analyzer door.
Press the BLUE button.
Insert Cal
Press Start

Table of Contents