65000159 Rev. 008 (06/17) 67
Calibration Reset
Calibration Reset Procedure
Action
Instrument
Information
Illustration
“Ready”
Place the Calibrator Card,
for the test type
calibration you want to
erase, on the door. Align
the pins on the door with
the holes in the Calibration
Card.
“Ready”
Close the analyzer door.
Press the BLUE button.
“Insert Cal –
Press Start”