68 65000159 Rev. 008 (06/17)
Calibration Reset Procedure
Action
Instrument
Information
Illustration
Press and hold the RED
button
Analyzer will beep four
times.
Release the RED button
when the beeping stops.
Example:
Erase
tPSA Cal?
If you do not want to erase
the calibration press the
RED button, again.
After “Erase tPSA Cal” is
displayed, the screen will
toggle between different
displays.
Press the BLUE button.
Example:
tPSA has
been erased
This will erase the
calibration.
Follow the Calibration
Procedure to re-calibrate
the analyzer.