EasyManua.ls Logo

Qualigen FastPack IP System - FastPack IP System Operation; Initial Checks and Sample Handling

Default Icon
87 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48 65000159 Rev. 008 (06/17)
FastPack
®
IP System Procedure
Action
Instrument
Information
Illustration
Verify that the analyzer is
in “Ready” mode.
Ready
Remove the FastPack
®
IP
from its box. Check to
make sure it is the correct
test type.
Ensure there are no
external fluid leaks or any
burst chambers on the
FastPack
®
IP.
Always keep FastPacks
®
stored in the refrigerator in
the upright position
indicated on the box, at 2-
8°C (36-46°F) prior to use.
FastPacks
®
can be taken
directly from the
refrigerator and used in
the analyzer.
Write the Patient’s
Name/ID # and the
operator’s initials on the
peel-off FastPack
®
IP label.
This information is
important for quality
control.
Closely follow FastSteps 2
through 5 from your
FastSteps Guide.
Use only Qualigen
recommended pipette tips.
The tips are available
through your local
distributor or Qualigen.
Be sure that the pipette
operates smoothly and
does not stick.
It is very important that
the tip is properly seated
on the pipette. If it is not,
there is a high likely hood
of short sampling which
will directly affect patient
results.

Table of Contents