65000159 Rev. 008 (06/17) 11
Calibrating the FastPack
®
IP System
When to Calibrate
Calibration ensures that each individual analyzer results are identical to all other analyzer results, and to
the master curve established by the manufacturer.
Additional information about calibration can also be found in the Direction Insert for each FastPack
®
IP
product as well as in the “A Word About Calibration” section found in the procedure manual.
What You Need
1. FastPack
®
IP System
2. FastPack
®
IP
3. FastPack
®
IP Calibration Kit
a. Calibration Card
b. Vial of Calibrator (1)
4. Pipette
* Use 25 µL pipette for hCG.
5. Pipette Tips (available from Qualigen)
6. Medical Gloves – used to avoid contact with
biohazard materials such as human blood, etc.
FOR FASTPACK
®
IP VITAMIN D USERS ONLY
7. Pretreatment Buffer