65000159 Rev. 008 (06/17) 69
Downloading Data
Smart Media Card Reader Equipped Model
The FastPack
®
IP System (when COM1 is configured for a computer) allows for the output of test results to
a personal computer (PC). The computer can be connected using a serial cable Qualigen Part Number
15000112. The cable connects to the RJ-45 jack on the rear panel of the FastPack Analyzer and terminates
with a female DB-9 connector compatible with input to a PC serial port. The cable is approximately 6 feet
in length. Data is automatically output at the completion of each test.
The output data is ASCII text with the following comma delimited fields:
Result Value (numerical or pass/fail for calibrator or controls)
Units
Date
Time
Instrument Serial Number
SW Revisions
Test Type
The communications protocol will be:
Serial Comm RS232
Baud Rate 9600
Parity None
Data bits 8
Stop bits 1
The “Recall last result” command can send the information for the last test run out to the PC.
Configuring Data Output Smart Media Card Reader Equipped Model:
The FastPack Analyzer has a setup menu that is accessed using the two buttons on the front panel. When
the instrument is in the standby mode “Ready” is displayed on the LED display. Pressing the red button
enters the setup menu. The red button scrolls through the choices, the blue button selects the choice
currently displayed.
Press the red button until the display shows one of the following:
Set to Printer” or “Set to Computer”
If the display shows “Set to Printer” it is currently configured for output to a computer and pressing the
blue button will change the configuration to make it work with the external printer. To leave the setup
menu with the FastPack Analyzer configured to output to a computer DO NOT PRESS THE BLUE
BUTTON, press the red button until the “Exit” option is displayed and then press the blue button to exit
the setup menu.
If the display shows “Set to Computer” it is currently configured for output to a printer and pressing the
blue button will change the configuration to make it work with a computer. To configure the FastPack
Analyzer for a Computer PRESS THE BLUE BUTTON. The system will beep, move the displayed text
from the first line of the display to the second line to confirm that the configuration was changed. There
will be a delay of a few seconds while the system stores the configuration in nonvolatile memory, then the
system will return to the standby mode and display “Ready”.
If the FastPack Analyzer is configured for output to an external printer, it will prompt the operator to
“Insert FastPack In Printer” each time a result is going to be sent out the communication port. The output
can be canceled by pressing the red button.