65000159 Rev. 008 (06/17) 13
Calibration Procedure
Action
Instrument
Information
Illustration
Open the analyzer door.
Remove the calibration
card.
Store card in Calibration
Kit.
Insert Cal
Press Start
Remove the FastPack
®
IP
from its box. Check to
make sure it is the correct
test type.
Always keep FastPack
®
IP
stored in the refrigerator,
in the position indicated on
the box, at 2-8°C (36-46°F)
prior to use. FastPacks
®
can
be taken directly from the
refrigerator and used in
the analyzer.
Write “CAL 1 or CAL 2 (or
CAL 3 for the Vitamin D
assay)along with the
operator’s initials on the
peel-off FastPack
®
IP label.
Invert the calibrator vial
6-8 times before opening.
Follow this procedure each
time you run a sample.