24 65000159 Rev. 008 (06/17)
Control Testing Procedure
Invert the buffer tube at
least 3 times to thoroughly
mix together the sample
and buffer. The act of
mixing the sample and
buffer releases the vitamin
D present, therefore
making it available for the
assay.
Follow the FastSteps for
Vitamin D Guide steps 9
and 10 to complete sample
delivery to the FastPack
®
IP.
Control Testing Procedure - Continued
Open analyzer door. Place
FastPack
®
IP on the door.
Align the pins on the door
with the holes in the
FastPack
®
IP.
It is only possible to align
the FastPack
®
IP in one
direction.
If the door is not closed,
the analyzer will not run.
The door will lock and the
analyzer will run the test.