CubiScan 125 Service Manual 23
Balancing the Load Cells
This section describes how to balance the CubiScan 125’s
load cells.
Items Needed
Phillips Screwdriver
Small Flathead Screwdriver
50 lb Weight (it is recommended that the maximum
weight is used)
Complete the following steps to balance the load cells.
Balanced load cells report the same weight. There are
three load cells located at each corner of the CubiScan
125. To balance the load cells the CubiScan 125 must be
on.
1. Make sure that there is no air blowing on the
CubiScan 125, as this can disrupt the balancing
process.
2. Zero the scale by tapping the
Zero button located on
the main screen.
3. On the CubiScan 125’s touchscreen, go to
Menu >
Diagnostics > Scale (1/2)
. This screen will display the
current weight in the
Wgt field.
4. Using the Phillips screwdriver, remove the four
screws holding the load cell summing board lid in
place and remove the lid. Refer to Figure 16.
5. To balance the load cells you will need to adjust the
variable resistors, which are also known as
potentiometers or POTs. To adjust a POT, use the
small flathead screwdriver to rotate the screw. See
Figure 17.
Rotating a POT clockwise will increase the added
Figure 16
Load Cell Summing Box
Figure 17
POTs
Chapter 5
Balancing the Load Cells
This section describes the procedure used to balance the load cells of the CubiScan 125.

Table of Contents

Other manuals for Quantronix CubiScan 125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantronix CubiScan 125 and is the answer not in the manual?

Quantronix CubiScan 125 Specifications

General IconGeneral
BrandQuantronix
ModelCubiScan 125
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals