R&S ZVA Performance Test
1177.5333.02 1.5 E-11
Checking the Maximum Output Power
Power sensor (item 3)
ZVA24, ZVA40, ZVA50, ZVA67, ZVB14, ZVB20 only: Calibration kit (item
4)
Connect power sensor to test port # of DUT, (ZVA24, ZVA40, ZVA50,
ZVA67, ZVB14, ZVB20 only: using the ff adapter from the calibration kit)
- [ PRESET ]
- [ SYSTEM CONFIG : Internal Reference ]
- [ SWEEP : Sweep Type : CW Mode ]
- [ POWER BW AVG : Power : 99 dBm ]
- [ MEAS : Wave Quantities: a1 Src Port 1
(a2 Src Port 2; More Wave Quantities: Wave Quantity: a3, Source
Port: Port 3…) ]
- [ START CENTER : CW Frequency : {f
Gen
*} ]
* For measurement frequencies, see Performance Test Report {f
GEN
}.
Read off the indicated power value.
Min output power .......................................... see Performance Test Report