Performance Test R&S
ZVA
1177.5333.02 1.12 E-11
Checking the Matching (raw)
Instrument:
Calibration kit (see Chapter “Test Equipment“, Item4)
Test cable (see Chapter “Test Equipment“, Item7)
R&S
®
ZVA settings:
- [ Preset ]
- [ Power : 0 dBm ]
1. Preparation/
test setup:
Connect the test cable to port x on the R&S
®
ZVA and perform a 1-port
calibration at the end of the cable.
Connect the end of the test cable to port # on the R&S
®
ZVA.
R&S
®
ZVA settings:
- [ Meas : Sxx]
- [ Marker ]
Measurement
Read off the network analyzer’s marker values (for marker frequencies
see Performance Test Report)