FSQ FVREF "Überschrift 2" \* MERGEFORMAT Fehler! Formatvorlage nicht definiert.
R&S ZVA Performance Test
1177.5333.02 1.11 E-12
Checking the Input Noise Level
Test equipment:
Calibration kit
(see Chapter “Test Equipment“, Item4)
Test setup:
Connect the Match Male from the calibration kit to port #
R&S
®
ZVA settings:
- [ Preset ]
- [ Meas : Wave Quantities: b# Src Port # ]
- [ System Config : RF Off Behaviour: Slow, Lower Broadband Noise ]
- [ Power : RF Off ]
- [ Meas Bandwidth : 10 Hz ]
- [ Marker ]
- [ Sweep : Sweep Type : CW Mode : CW Frequency : {f
GEN
*}]
- Service Function 1.0.0.1.1 (see chapter 3 ‘Service Functions’, Service
Level 2)
- [ Measure : Wave Quantities : More Wave Quantities… : Properties :
Detector : RMS : Meas. Time: 500ms ]
- [ Sweep : Single : Restart]
* For measurement frequencies, see Performance Test Report {fGEN}.
Measurement:
Read off the noise level indicated by the markers on the DUT.