Repairs R&S ZVA
1177.5333.02 3.66 E-11
Replacing the Reflectometers on ZVA80-BU
Port1 and port2: See Replacing a Reflectometer RM 40 and RM50
Port3 and port4:
(See Chapter 5, drawing 1312.6508)
(See Chapter 5, Spare Parts List, Item 100, and drawing 1166.1660)
The boards are located under the motherboard.
Opening the instrument and removing the board
Turn off the instrument and disconnect from the mains, screw off the 4 rear-panel feet (460) and
four screws (455), remove the two side handles (450) and pull off the enclosure (400) backwards.
Remove the 4 screws in the front handles (410), right and left, and take off the front handles.
Remove the countersunk screw (315) and pull off the front cover (300) forwards
Remove the bottom instrument cover (298) after undoing the 18 countersunk screws (299).
Remove RF connection cables (ZVA50-B16)
Disconnect the source cable, the LO cable, the two IF cables and the 50-pin control cable from the
reflectometer.
Remove the 4 screws (3132) and the 2 screws (3133) in the front of the instrument next to the port
connector.
Remove the 5 screws (3131) and take out the reflectometer
Installing the board and reassembling the instrument
Insert the new board into the instrument and reconnect any cables that have been disconnected.
N.B.: Use the lettering on the motherboard as an aid.
Screw in the 6 screws (3132, 3133) in the front of the instrument next to the port connector.
Screw in the 5 screws (3131).
Refit support plates (3140, 3142) and reflectometers A530, A540, A560 (only with option ZVT20-
B63, -B64, -B66).
Fit the instrument base cover (298) and secure with 18 countersunk screws (299).
Replace the front cover (300) and secure with the countersunk screw (315).
Refit the 2 front handles (410) using 4 screws.
Slide the enclosure (400) back on and screw the 4 rear-panel feet (460) and the 2 side handles
(450) into position.
Connect the mains cable and turn on at the mains switch. The instrument is now in the standby
mode
Select Service Level 2 (see Service Functions).
When the instrument has been started, check the protocol file for errors:
[ INFO : Error Log ]
Record the generator and receiver correction data (see Recording Correction Values).
Perform the factory system error calibration (see Factory System Error Calibration).