Repairs R&S ZVA
1177.5333.02 3.58 E-11
Replacing Generator Board (ZVA8 only)
(See Chapter 5, Spare Parts List, Item 110 and drawing 1302.4960.08)
Opening the instrument and removing the board
Remove the reflectometer (see chapter Replacing Reflectometer Board A510 to A540)
Remove the cover (130)
Remove the fan cable
Pull off the generator board (110)
Installing the board and reassembling the instrument
Fit the new generator board (110)
Reconnect the fan cable
Fit the cover (130)
Reinstall the reflectometer (see chapter Replacing Reflectometer Board A510 to A540)
Connect the mains cable, turn on at the mains switch and press the ON key.
Select Service Level 2 (see Service Functions).
When the instrument has been started, check the protocol file for errors:
[ INFO : Error Log ]
Record the generator and receiver correction data (see Recording Correction Values in chapter 2).
Perform the factory system error calibration (see Factory System Error Calibration in chapter 2).