FSQ FVREF "Überschrift 2" \* MERGEFORMAT Fehler! Formatvorlage nicht definiert.
R&S ZVA Performance Test
1177.5333.02 1.15 E-11
Checking the DC Measurement Inputs
DC Power Supply NGSM 32/10 (see Chapter “Test Equipment“, Item 10)
Multimeter URE3 (see Chapter “Test Equipment“, Item11)
DC cable (see Chapter “Test Equipment“, Item12)
Connect the Power Supply to the Input DC MEAS 1V (DC MEAS 10V) of
the ZVA using the DC cable
- [ Preset ]
For DC Meas 1 V:
- [ Meas : More : DC Inputs : DC Meas 1 V]
- [ Format : Real]
- [ Scale : Scale/Div : .25 x1]
- [ Marker ]
For DC Meas 10 V:
- [ Meas : More : DC Inputs : DC Meas 10 V]
- [ Format : Real]
- [ Scale : Scale/Div : 2.5 x1]
- [ Marker ]
Set Power Supply to DC values U
DC
using Multimeter URE (DC values
see Performance Test Report) and connect it to pos. and neg. input
Read off the DC level U
DC ZVA
indicated by the marker.
Deviation = U
DC
- U
DC ZVA