R&S ZVA Performance Test
1177.5333.02 1.9 E-11
Checking the Harmonics
Test cable (see Chapter “Test Equipment", Item7)
20dB attenuator (see Chapter “Test Equipment", Item7)
Connect the attenuator to one of the other ports (portx)
Connect one end of the test cable to the attenuator
Perform a power calibration (with test cable and attenuator, for power
settings see Performance Test Report).
If there are different power settings, depending on the frequency range,
for each power setting a calibration must be performed.
- [ Preset ]
With option K4:
- [ Mode : Harmonics : 2
nd
Harmonic (Port# to Portx) : Relative
Measurement ]
- [ Mode : Harmonics : 3
nd
Harmonic (Port# to Portx) : Relative
Measurement ]
Without option K4:
- Service Function 1.0.0.0.1 (see chapter 3 ‘Service Functions’, Service
Level 2)
- [ Stop : fmax/2] (2
nd
Harmonic)
- [ Stop : fmax/3] (3
nd
Harmonic)
- [ Pwr BW AVG : {p
GEN
*}]
- [ Marker : {f
GEN
*} ]
* For marker frequencies and power settings, see Performance Test
Report.
Read off the marker values