1177.5333.02 3.81 E-11
Troubleshooting Boot Problems
Fault: ZVA does not start the measurement application.
The first action the ZVA performs after power-up is booting BIOS for the processor. When the processor
has been successfully initialised, the Windows XP start-up procedure begins. Then, the measurement
application is loaded as an autostart program.
If there are errors during the boot phase, messages indicate possible defects.
The message “No System Disk or Disk error…” indicates that the hard disk data is corrupt. If this is the
case, replace the hard disk.
If the operating system on the hard disk has been corrupted and so cannot be loaded correctly,
Windows XP outputs a “blue screen”.