Performance Test  R&S
 
ZVA 
1177.5333.02  1.6  E-11 
Checking the Accuracy of Output Power  
 
Power sensor or 
Power meter with power sensor 
(see Chapter “Test Equipment“, Item2) 
 
Connect the power sensor to port # 
 
For measurement frequencies, see Performance Test Report {f
GEN
*}. 
 
- [ Preset ] 
- [ Meas :Wave Quantities: a# Src Port # ] 
- [ Sweep : Sweep Type : CW Mode :  
CW Frequency : {f
GEN
*} ;  Power : -10 dBm] 
- [ Mode : ALC : On] 
- [ Sweep : Single : Restart] 
 
* For measurement frequencies, see Performance Test Report {f
GEN
}.  
 
Read off the levels indicated by the power sensor 
 
Level deviation = L
SENSOR
 – (-10 dBm)