R&S ZVA  Repairs 
1177.5333.02  3.33  E-11 
Putting into operation 
 
Firmware 
  Perform an Analyzer BIOS update for FMR9 (see https://gloris.rohdeschwarz.com, Document link: 
Firm- \ Software) 
  Perform a HD recovery using ZVABT Recovery CD FMR9 
After reboot, switch analyzer application screen to windows screen. 
Use windows explorer to browse and search the batch file to configure the display type. 
Select and execute C:\Display\Tools\image_set_601.bat. 
Reboot 
  Perform a firmware update. 
  Download from http://www.rsd.de/  
Select Downloads  Select your product: "ZVA" and mark "Firmware"  Go 
Firmware Version 2.81 or higher is necessary.  
Netcon PCINT FPGA Version 8.5 or higher is necessary. 
 
Restoration of Factory settings 
  Switch the unit on by pressing the “Standby” button in front. 
  Copy the stored files “Factory.cal” and “Factory.dat” from USB drive to 
C:\Documents and Settings\All Users\Application 
Data\Rohde&Schwarz\Nwa\Data\calibration 
  Switch the unit off and on by pressing the “Standby” button in front. 
  Select Service Level 2 (see chapter Service Functions). 
  When the instrument has been started, check the protocol file for errors: 
[ INFO : Error Log ]