R&S ZVA Repairs
1177.5333.02 3.33 E-11
Putting into operation
Firmware
Perform an Analyzer BIOS update for FMR9 (see https://gloris.rohdeschwarz.com, Document link:
Firm- \ Software)
Perform a HD recovery using ZVABT Recovery CD FMR9
After reboot, switch analyzer application screen to windows screen.
Use windows explorer to browse and search the batch file to configure the display type.
Select and execute C:\Display\Tools\image_set_601.bat.
Reboot
Perform a firmware update.
Download from http://www.rsd.de/
Select Downloads Select your product: "ZVA" and mark "Firmware" Go
Firmware Version 2.81 or higher is necessary.
Netcon PCINT FPGA Version 8.5 or higher is necessary.
Restoration of Factory settings
Switch the unit on by pressing the “Standby” button in front.
Copy the stored files “Factory.cal” and “Factory.dat” from USB drive to
C:\Documents and Settings\All Users\Application
Data\Rohde&Schwarz\Nwa\Data\calibration
Switch the unit off and on by pressing the “Standby button in front.
Select Service Level 2 (see chapter Service Functions).
When the instrument has been started, check the protocol file for errors:
[ INFO : Error Log ]