Repairs R&S ZVA
1177.5333.02 3.96 E-11
Testing the Synthesizer Board
N.B.: Synthesizers with part numbers 1145.xxxx or 1300.xxxx are no longer available. In case of
defect please contact the Central Service in Munich.
Disconnect source (or local) cable at the reflectometer (1 to 4, depending on which synthesizer
section is to be tested).
Connect the end of the cable to the spectrum analyzer input via an adapter cable and adapter
SMA-N.
N.B.: Bend the source cable as little as possible, if necessary secure adapter cable with
adhesive tape.
Enter service function 2.21.2.18.17512345 (IF = 17.512345 MHz).
[ Mode : Spurious Avoidance : LO > RF ]
Set the frequencies listed in the table on the ZVA and check the values.
[ Sweep : Sweep Type : CW Mode : CW Frequency : Hz]
[ Sweep : Single : Restart]
Synthesizer-LS 1305.5300.xx
Source Mod. 02 and Mod. 20
Frequency
(ZVA setting)
Source Level
(Mod. 02)
2nd harmonic
3rd harmonic
300 kHz
+3 dBm to -5 dBm
< -27 dBc
< -30 dBc
100 MHz
+3 dBm to -5 dBm
< -27 dBc
< -30 dBc
1 GHz
+3 dBm to -5 dBm
< -27 dBc
< -30 dBc
2 GHz
+3 dBm to -5 dBm
< -27 dBc
< -30 dBc
3 GHz
+3 dBm to -5 dBm
< -20 dBc
< -30 dBc
4 GHz
+3 dBm to -5 dBm
< -20 dBc
< -30 dBc
6 GHz
+3 dBm to -7 dBm
< -27 dBc
< -30 dBc
8 GHz
+3 dBm to -7 dBm
< -27 dBc
< -30 dBc
Isolation of source switch1 to 4
ZVA frequency 8 GHz
Port no.
Port setting
Source output
Isolation
1
Port1 inactive
Port2 active
X168
X169
-90 dBc ±5 dB
2
Port1 active
Port2 inactive
X168
X169
-90 dBc ±5 dB

Table of Contents