Repairs R&S ZVA
1177.5333.02 3.110 E-11
Testing the Reflectometer RM44 and RM50
Generator levels
It is assumed that the synthesizer section (synthesizer1) associated with the reflectometer to be tested
is OK.
Loosen cable W514A (GEN -> coupler unit) or W512 (GEN -> attenuator) at both ends and screw
off at the generator output GEN.
Connect the generator output to the spectrum analyzer using the K-cable (ZVA40) or V-cable
(ZVA50).
Set the ZVA to the CW sweep mode.
Set the frequencies listed in the table.
Set the power listed in the table.
Set power to
(without
GenAtt)
Set power to
(with GenAtt)
With the stated levels, the cable loss must still be taken into account.
Receiver levels
The following is assumed:
The LO synthesizer section (synthesizer1 mod. 20, synthesizer2) associated with the
reflectometer to be tested is OK.
The network controller associated with the reflectometer to be tested is OK.
One reflectometer in the instrument is functioning.