Equator 300 machine specifications
Comparison uncertainty
± 0.002 mm
Scale resolution
0.0002 mm
Maximum move velocity
500 mm/s
Maximum move acceleration
2500 mm/s
2
Storage temperature
-25 °C to +70 °C
Operating temperature
+5 °C to +50 °C
Operating humidity
Maximum 80 %RH at 40 °C, non-condensing
Machine electrical supply requirements
24 V DC supplied direct from Equator Controller
Repeatability of piece part in fixture
1 mm
Maximum workpiece weight (including fixture plate)
25 kg
Protective equipment suggested*
Eye protection and safety shoes
Guards
None
The process of measuring on an Equator involves defining a series of gauge points on the component
surface. Periodic calibration of a master part on a CMM establishes datum values for each gauge point. The
same gauge points on the same master part are measured on Equator (known as ‘mastering’) to establish
a correlation with the certified CMM. Subsequently, a regular ‘re-mastering’ process is used to account for
changing environmental conditions.
Size and position measurements made immediately following re-mastering will have a comparison uncertainty
of ±0.002 mm relative to the certified measurements of the master part. This specification applies where each
part is fixtured to within 1 mm relative to the master part.
*Customers should complete their own risk assessment on delivery of the machine to define their own PPE
requirements.
Equator 300 machine specifications - Scanning
Probe type
Renishaw 3 axis SP25 analogue scanning probe
Maximum scanning velocity*
200 mm/s
Maximum scanning acceleration
1500 mm/s
2
Maximum scanning rate
1000 points/s
Equator 300 machine specifications - Touch Trigger
Probe type
Renishaw 3 axis TP20 kinematic touch-trigger probe
Maximum touch point velocity
10 mm/s
Maximum touch acceleration
1500 mm/s
2
Equator 300 machine labels
A warning label on each of the three support arms for the constraint mechanism warns
not to lift the machine by the support arms. The Equator machine should only be lifted
using the four hand holds on the base casting.
* Maximum scanning speed cannot be guaranteed as it is affected by feature type/size/finish and stylus
configurations.
9