6.5
Fault-finding
Symptom Cause Action
Probe triggers but
the OMI-2T does not
respond.
3 second switch-on mode
selected.
Probe out of range.
Transmission beam obstructed.
Probe set to legacy transmission.
Recongure to the standard
switch-on delay.
Review performance envelopes.
Check the OMP400 and receiver
windows are clean, and remove
any obstruction.
Recongure to modulated
transmission.