Accessories
36
To attach a bacterial/viral filter:
1. Fit the bacterial/viral filter to the inspiratory port of the device.
2. Connect the air tubing to the other side of the filter.
3. Perform the Learn Circuit function.
4. Attach the patient interface to the free end of the air tubing.
WARNING
To prevent the risk of cross-contamination, a bacterial/viral filter is mandatory if the device is to
be used on multiple patients.
The expiratory module, internal bacterial/viral filter, expiratory flow sensor and blue membrane
come into contact with exhaled gases but do not form part of the inspiratory pathway.
Adding supplemental oxygen
Oxygen may be prescribed by your clinician.
The Astral device is designed to be compatible with levels of supplemental oxygen up to 30 L/min.
At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration will vary depending on the
Ventilation mode and settings, patient breathing pattern, mask selection, and leak rate.

Table of Contents

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery
Pressure Support0 to 30 cm H2O