Detecting circuit disconnection and de-cannulation
58
Detecting circuit disconnection and de-cannulation
Inadvertent disconnection of a circuit component or accidental removal of a cannula poses a hazard to a
dependent patient. Astral is equipped with a number of alarms that when used in conjunction with the
Disconnection Alarm are able to reliably detect circuit disconnection (including de-cannulation).
The optimal alarm may depend on the therapy target and circuit type as shown in the table below.
However, Astral provides a number of alarms that can be configured by your clinician specifically for this
purpose.
WARNING
Alarm settings are sensitive to any changes to the circuit, ventilation settings or co-therapy. Test
the effectiveness of the alarm after any of these changes are made.
CAUTION
Alarms should be configured and tested to ensure that circuit disconnection and de-cannulation is
detected. We recommend configuring and testing multiple alarms and testing disconnection at the
ventilator and at the cannula. Independent monitoring can be used as an alternative.
The following table provides the most appropriate alarms for use in detecting circuit disconnection.
Pressure target modes Volume target modes
Single with leak
Disconnection alarm
Low pressure alarm
Low Vte alarm
Low MVe alarm
Apnoea alarm
Leak alarm
SpO
2
alarm
N/A
Single with valve
Disconnection alarm
Low pressure alarm
Low Peep alarm
High Vti alarm
High MVi alarm
Apnoea alarm
SpO
2
alarm
Disconnection alarm
Low pressure alarm
Low PEEP alarm
Apnoea alarm
SpO
2
alarm
Double with valve
Disconnection alarm
Low pressure alarm
Low Vte alarm
Low MVe alarm
Apnoea alarm
Leak alarm
SpO
2
alarm
Mouthpiece
Disconnection alarm
Low pressure alarm
High Vti alarm
High MVi alarm
Apnoea alarm
SpO
2
alarm
Disconnection alarm
Low pressure alarm
Apnoea alarm
SpO
2
alarm

Table of Contents

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery
Pressure Support0 to 30 cm H2O

Related product manuals