Accessories
42
Power management
Helpful hints!
Connect the ventilator to the mains power whenever possible. In the event of battery failure, connect
to mains power immediately to resume ventilation.
An external power source (Astral External Battery or RPSII) should always be available for
ventilator-dependent patients.
An external power source (Astral External Battery or RPSII) should be in use in mobile situations,
including when mains power is unavailable or disrupted. Do not rely solely on the internal battery for
mobile use.
Ensure the external battery is sufficiently charged before using in mobile situations.
WARNING
Beware of electrocution. Do not immerse the device, power supply or power cord in water.
Make sure the power cord and plug are in good condition and the equipment is not damaged.
Keep the power cord away from hot surfaces.
Explosion hazarddo not use in the vicinity of flammable anaesthetics.
The Astral device can be used with different power sources:
Mains power
Astral External Battery
ResMed Power Station II
External DC power supply (eg, car 12V power outlet)
Internal battery
For information on power supplies and sources, refer to the Technical Specifications (see page 73).

Table of Contents

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery