Assembling patient circuits
English 29
Connecting a mouthpiece circuit
The mouthpiece circuit is a single limb circuit with no expiratory valve or intentional leak. This circuit is
not intended to support continuous exhalation into the circuit. For patients that may prefer continuous
exhalation into the circuit, a circuit with expiratory valve or intentional leak should be considered.
To connect a mouthpiece circuit:
1. Check the device is fitted with a single limb leak adapter. Otherwise, change the adapter.
Note: The Astral 150 can also support mouthpiece circuit using a double limb adapter.
2. Connect the inspiratory limb to the inspiratory port.
3. Attach any required circuit accessories (eg, filter).
4. Select the circuit type and perform a Learn Circuit.
5. Attach the patient interface (eg, mouthpiece) to the free end of the air tubing as appropriate.

Table of Contents

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery
Pressure Support0 to 30 cm H2O