Alarms
English 51
Indicator Description
4
Alarm mute/reset button State:
no light no active alarms
steady light active alarm/s
flashing light alarm mute on.
This button also allows you to:
mute the audible alert
reset the currently displayed alarm (if permitted).
5
Alarm bar Indicates the priority of the alarm in the Alarm display.
Alarm priority
Alarms are classified into relative priority (high, medium and low) according to the severity and urgency
of the alarm condition. Respond to all alarms. An immediate response is required for high priority alarms.
Alarm priority
Alarm bar Audible alert
High
Red flashing light 10 beeps every 5 seconds
Medium
Yellow flashing light 3 beeps every 15 seconds
Low
Yellow steady 2 beeps every 25 seconds
Helpful hint!
For suggestions on resolving most common alarms, refer to Alarms Troubleshooting.

Table of Contents

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery
Pressure Support0 to 30 cm H2O

Related product manuals