Packaging, Transport and Installation
16
NOTICE
N7.0021
Ambient temperature
Temperatures outside the permitted range
Electronic and mechanical components may be damaged.
The performance data alter to an unknown extent.
Do not exceed or fall below the permitted temperature range (5 °C to
40 °C ambient temperature) of the device.
Maximum relative humidity < 80 % (at ambient temperatures 31 °C)
For ambient temperatures U
T
between 31 °C and 40 °C, the maximum relative humidity value L
F
linearly decreases according to L
F
= (U
T
55) / 0.3:
Ambient temperature
Max. rel. humidity
31 °C
80 %
33 °C
73.3 %
35 °C
66.7 %
37 °C
60 %
39 °C
53.3 %
40 °C
50 %
NOTICE
N8.0015
Humidity
High relative humidity
Electronic and mechanical components may be damaged.
The performance data alter to an unknown extent.
The relative humidity in the vicinity of the device should be kept as
low as possible.
4.5 Electrical Connection
WARNING When connecting the power cable to the mains supply, use an external fuse that
complies with the regulations applicable to the place of installation.
Check the type plate for details on the necessary voltage and frequency for the device.
The listed values must agree with the existing mains supply.
Only use the supplied power cable to connect the device to the mains supply.
NOTICE
N9.0022
Electrical connection
Failure to observe the values on the type plate
Electronic and mechanical components may be damaged.
Connect the device only to mains supply matching the values on the
type plate.