The Rotor Beater Mill of the Retsch GmbH is universally deployed in almost all areas of industry
and research, from sample preparation in the laboratory to larger sample volumes in a technical
centre or plant.
Only grinding sets from Retsch GmbH may be used. This device is generally not designed for
the grinding of wet or moist materials.
Handling of food, pharmaceutical and cosmetic products
Processed products
Food, pharmaceutical and cosmetic products, which were processed with
the device must not be consumed, used or circulated.
Dispose these substances in accordance with the applicable
regulations.
Range of application of the device
Long-term operation
This laboratory device is designed for eight-hour single-shift operation
with a duty cycle of 30 %.
This device may not be used as a production machine nor is it
intended for continuous operation.
6.2 Principle of Operation
The grinding and deagglomeration in the SR 300 is achieved by impact, collision and shearing
effects. The feed material passes through the feed hopper into the grinding chamber where it is
comminuted between the rotor and the selected ring sieve or sieve and grinding insert. As soon
as the grinding material is smaller than the mesh size of the sieve, it enters the collecting
receptacle.