8
GA--A 181
3Setup
W ARNING!
This device is not protected against explosions.
Explosion hazard.
Do not operate this device in areas where there is a danger of
explosions.
CAUTION!
Danger of faults and malfunctions.
To guarantee the safety of the user, the patient and others use only
accessories and spare parts specified by the manufacturer of this
product.
Other accessories or spare parts can cause the emission of
increased electromagnetic radiation or reduced immunity against
interference.
.
IMPORTANT!
Medical devices are subject to special precautions with regard to electro-
magnetic compatibility (EMC).
Make sure you observe the notes on EMC for installation and operation.
Medical electrical devices can be influenced by mobile HF communica-
tion devices.
If it is necessary to stack the devices or place them next to each other
and HF interference is observed, make sure you observe the intended
use of the devices.
CAUTION!
Danger of infection from unsterile accessories.
Sterilize reprocessed parts and accessory items before use.
Follow the sterilization instructions for the product in question!
.
IMPORTANT!
Make sure that the mains/grid voltage and the voltage indicated on the
identification plate are the same. Connect the device only through the
supplied power cord or a power cord meeting the same specifications.
.
NOTE!
During operation place the device on an even, level, non--slippery
surface.