7.2 Reprocessing of accessories
CAUTION!
Danger of infection due to unsterile parts and accessories.
Sterilize reprocessible parts and accessories before use. Follow the
regulations on sterilization valid in your country.
.
IMPORTANT!
Follow the procedures described in the “General instructions and
notes on the reprocessing of R. Wolf products, accessories and
devices” (order no.: GA--J 020) .
7.2.1 Wet preparation at the point of use
After use immerse used accessory items such as transducer, sonotrodes
and suction tube in a disinfectant solution for wet preparation at the point
of use.
For this pupose follow the disinfectant manufacturer s instructions!
Z Disconnect connection bar (17) of the transducer (15) from the device.
Disconnect suction tube (19) from the tube connector on the device.
Immerse accessories in the solution for wet preparation at the point of
use.
7.2.2 Disassembly before cleaning
Remove accessories items before cleaning.
Z Unscrew the sonotrode (16) with the supplied open--ended wrench
(size: quarter inch) by turning counter clockwise (right hand thread).
Disconnect suction tube (19) from tube connector (18) of the trans-
ducer .