26
GA--A 181
7.2.5 Disinfection
For disinfection, immerse accessories in disinfectant solution. Before you
immerse the parts fill the channels of the transducer, sonotrode and suc-
tion tube completly with disinfectant using a syringe.
Follow the disinfectant manufacturer’s instructions!
After disinfection dry the channels of the accessory items with com-
pressed air and the surfaces with a sterile cloth.
.
NOTE!
Do not use disinfectants containing chlorine or phenole derivate for the
disinfection of R. Wolf products.
Avoid immersion/soaking times of more than two hours!
7.2.6 Steam sterilization
Never steam--sterilize the tranducer and the sonotrodes in assembled
condition!
To sterilize the transducer, we recommend the transducer reprocessing-
basket (38011.501). See instruction manual GA--J 040.
Z Steam sterilization at 134°C (272°F) using the fractional method.
7.2.7 Assembly
Assemble accessories only immediately before use and in reverse order
as described under ’Disassembly’.
Follow assembly instructions for sonotrodes under section 5.
Perform the neccessary checks before each use in accordance with
section 4.
7.2.8 Reprocessing of foot switch
The device can be cleaned with a soft cloth moistened with surface disin-
fectant, alcohol or spirit.
Follow the disinfectant manufacturer’s instructions!