EasyManua.ls Logo

Richard Wolf 2271 - Technical Data of Transducer (2271.501); Technical Data of Footswitch (2030.12); Operating, Storage, Transport and Shipping Conditions

Richard Wolf 2271
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32
GA--A 181
8.2.4 Technical data of transducer (2271.501)
Protection against electric shock BF type applied part
Degree of protection against the ingression of liquids IP 67
Degree of protection in the presence of flammable
gasses
This device is not protected against explosions.
(Do not operate this device in flammable environments)
Weight 0.300 kg (0.7 lbs)
Dimensions (dia. x length) 30 mm x 143 mm
8.2.5 Technical data of footswitch (2030.12)
Degree of protection against the ingression of liquid: IP 68
Degree of protection in the presence of flammable
gasses
This device is not protected against explosions.
(Do not operate this device in flammable environments)
Weight 0.730 kg (1.61 lbs)
Dimensions (Width x height x length) 70 mm x 61 mm x 220 mm
8.3 Operating, storag e, transport and shipping conditions
Operating conditions
ambient temperature + 10°Cto+40°C,
30% to 75% rel. humidity,
atmospheric pressure 700 hPa to 1060 hPa
Storage, transport and shipping conditions
ambient temperature -- 20°Cto+60°C,
10% to 90% rel. humidity,
atmospheric pressure 700 hPa to 1060 hPa
.
NOTE!
To avoid damage to the products during transport or shipment we recommend using the original pack-
aging material.