EasyManua.ls Logo

Rigol DG1022

Rigol DG1022
147 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
© 2008 RIGOL Technologies, Inc.
Users Guide for DG1022
2-70
Test/Cal Settings
Press Utility Test/Cal, enter the following menu.
Figure 2-78 Test Menu
Table 2-30 Test Setting Menu
1. Test
Press Utility Test/Cal Test, to perform self-testing.
When the instrument is powered, the system will perform the self-testing
automatically. This limited test can test if the instrument could work ordinarily.
To perform a general test needs 3 seconds. If all the tests have been passed, you
can use the generator freely.
If the general test has been passed, the screen will show “Self-Test Passed”; if not,
it will also tell you self-test failed and display the error message.
Menu
Settings
Explanation
Test
Perform System Self-Test
Info
Show the system information
PassWd
Input the password for Secure calibration
SecOn
SecOff
On
Turn On Secure Setting, disables the
calibration by anyone
Off
Turn Off Secure Setting, enables the
calibration by anyone
Cal
Enter Calibration interface

Table of Contents

Other manuals for Rigol DG1022

Related product manuals