Troubleshooting RIGOL
User’s Guide for DSA1000 Series
the error.
This code indicates that there is no room in the queue and an
error occurred but was not recorded.
-360 Communication error.
-365 Time out error.
Time out when communicating with USB-GPIB Switching Module,
please restart the module.
300 Print paper error.
Failure to print. Please check whether the printer paper is available and
in the correct position.
301 Print ink error.
Indicates that an error occurred when printing. Please check the
quantity of ink or confirm whether the ink is installed correctly.
302 Print hardware error.
Indicates that a hardware error has occurred. Please check it.
303 Print file type error.
Failure to print owing to wrong print file type.
304 Unknown print error.
310 Updating error with wrong image file.
An error occurred when updating firmware owing to a wrong format of
the image file or the file has already been damaged.
311 Updating error with wrong revision.
An error occurred when updating firmware because of a revision
incompatibility.
312 Updating error with failing to write flash.
An error occurred when updating firmware owing to a failure to write to
the flash.
The file path received may not exist or an illegal path format was
received.
321 Invalid input string.
An illegal string has been received.
The file name is too long to be accepted. The length should not exceed
48 characters.
323 USB-Disk failed to install.
The USB storage device can not be correctly installed, please examine it
for possible damage. Any question, please contact your distributor or
technical support.